The current CLIA total allowable error (TEa) limits by specialty: chemistry, hematology, toxicology, and immunology, with 42 CFR §493 citations.
Frequently Asked Questions
Does CLIA require labs to use CLIA TEa for calibration verification, or can we use manufacturer criteria?
No. CLIA requires calibration verification to be performed and documented and leaves the acceptable limits to the laboratory (42 CFR §493.1253(b)(3) and §493.1255(b)(2)); it does not specify numeric criteria. A medical director or designee may adopt the §493 PT limit for the same analyte as the criterion and should document why it fits the clinical need. CMS said in the preamble to CMS-3355-F (87 FR 41194, July 11, 2022), which is not regulation text, that PT limits were written to grade proficiency testing and were not intended as performance goals, and that goals for accuracy and precision should rest on clinical needs and the manufacturer's labeling. §493.1255(b)(1) also has the laboratory follow the manufacturer's calibration verification instructions.
What's the difference between TEa used for proficiency testing and TEa used for calibration verification?
The table in 42 CFR Part 493, Subpart I applies to proficiency testing: it defines whether your PT results are graded as acceptable. For calibration verification, §493.1255(b)(2) has the laboratory set its own acceptable limits. A medical director or designee may adopt a PT limit as that criterion and should document why it fits the clinical need, but the CFR does not give that choice the force of the PT table.
Our analyte isn't in the CLIA TEa table. What do we use?
If an analyte isn't regulated under CLIA proficiency testing requirements, Subpart I publishes no PT limit for it. In that case, labs should use manufacturer-stated allowable error, biological variation-based goals (from EFLM or RCPA tables), or medical decision-based criteria established by the laboratory director or designee. Document the rationale for whatever criteria you choose.
How often does CLIA TEa change?
Infrequently, but the CLIA proficiency testing final rule (CMS-3355-F, published July 11, 2022) was the first major update in decades for many analytes. Before that, some values dated to 1992. The updated Subpart I criteria took effect July 11, 2024. Check the current eCFR rather than relying on reference cards or QC software that may not have been updated.
What does 'target value ±X% or ±Y units, whichever is greater' mean in practice?
The dual criterion accounts for concentration-dependent error. At low concentrations, a fixed percentage can be smaller than what's analytically meaningful, so the absolute value floor protects the criterion from being trivially easy to pass near zero. Apply both. The one that allows a larger absolute difference is the governing criterion at that concentration.