How to interpret EP15-A3 precision verification reports for CLIA compliance: SD, CV, total imprecision, and what surveyors look for in your documentation.
Frequently Asked Questions
Why is precision verification required?
42 CFR §493.1253(b)(1) requires a laboratory introducing an unmodified, FDA-cleared or approved test system to demonstrate, before reporting patient results, that it can obtain precision comparable to the manufacturer's. A modified system must have its precision established under §493.1253(b)(2).
What do mean, SD, and CV mean on the report?
Mean is the average of replicates; standard deviation is how widely replicates spread around it; coefficient of variation is the SD as a percent of the mean, CV = (SD / mean) x 100.
How many replicates does it need?
CLSI EP15-A3 uses a 5-day by 5-replicate design at two or more concentrations. With fewer replicates, the confidence interval around SD widens and the estimate is less trustworthy.
What do Pass, Fail, and Uncertain mean?
Against a vendor SD goal, Pass means the upper bound of the 95 percent confidence interval for the observed SD is below the goal; Fail means the lower bound exceeds it; Uncertain means the goal falls inside the interval and more replicates may be warranted.
What acceptance criterion should I use?
For precision, the regulation's comparison is the manufacturer's claim: 42 CFR §493.1253(b)(1)(i) has the laboratory show precision comparable to the manufacturer's, so the manufacturer's published within-run SD is the starting point. A medical director or designee may adopt another criterion, such as a fraction of a CLIA PT limit, and should document why it fits the clinical need. Final acceptance rests with the medical director or designee.